Precise measurement of Particulate Matter and CO2 for expert environments.
PMsenseCR is a PM and optionally CO2 transmitter, ideal for pharmaceutical and high-tech cleanrooms where contamination control is critical.
It ensures compliance with ISO 14644 standards using laser scattering technology to detect and classify particles by size.
Reliable & Maintenance-Free
Built for long-term stability, PMsenseCR offers fast response, high sensitivity, and extended operating life – no routine maintenance required.
Optional Sensing
Add an integrated sensor for comprehensive indoor monitoring.
Flexible Operation & Output
Choose between continuous or cyclic measurement modes to optimize sensor lifespan. Supports RS485 MODBUS-RTU/ASCII and 4…29 mA outputs, with five measurement channels:
| Power supply: | 7…30 Vdc (15…30 Vdc for 0…10 V analog outputs) |
|---|---|
| Power consumption: | 25 mA @ 24 Vdc during measurement 4 mA in stand-by (only for cyclic operating mode) The indicated consumption does not include the consumption due to the analog outputs |
| Weight: | 330 g |
| Dimensions: | 120 x 94 x 71 (excluding M12 connector) |
| Connection: | M12 8-pole circular connector |
| Operating conditions: | -20…+70 °C / 0…95 %RH / 500…1500 hPa |
| Measuring range: | < 3.3 x 109 pcs/m3 |
| Accuracy: | ±(50 ppm+3% of measurement) @ 25 °C and 1013 hPa |