Temperature & Testing Data
This data is calculated with the transducer, wedge, and asset all at 150°C (peek data) or 200°C (celazole data). During actual inspections, there will be a temperature gradient from the asset up to the transducer.
High-Temp Wedges – Celazole U-60
Peek | |||||||
---|---|---|---|---|---|---|---|
Temp (°C) | 1st BW Gain (dB) Low / Low | Velocity (in./uS) | % Dif V from 25 °C | Gain at 80% FSH | % Dif BW from 25(°C) | Fc (MHz) | % Dif Fc from 25 (°C) |
25 | 10 | 0.117 | 0 | 16 | 0 | 4.88 | 0 |
50 | 10 | 0.116 | 0.77 | 16 | 0 | 4.8 | 1.64 |
75 | 10 | 0.116 | 1.20< | 16.5 | 3.13< | 4.76 | 2.46> |
100 | 12.5 | 0.115 | 1.97< | 17.5 | 9.38 | 4.66 | >4.51 |
120 | 15 | 0.114 | 2.65 | 19.5 | 21.88 | 4.27 | 12.50 |
150 | 20 | 0.113 | 3.68 | 23.5 | 46.88 | 3.86 | 20.90 |
175 | 26 | 0.113 | 3.42 | 29.5 | 84.38 | 3.91 | 19.88 |
200 | 39 | 0.112 | 4.53 | 39 | 143.75 | 6.05 | -23.98 |