High-Temp Dual Linear: Dual-Linear Corrosion Array


 
The High-Temp Dual-Linear Corrosion Array is optimized for corrosion and erosion inspection at elevated temperatures. The transducer and its replaceable delay line is designed to withstand temperatures up to 200°C (392°F). This dual array features 32 transmit and 32 receive elements to provide larger beam coverage than conventional dual-element transducers. The transmit and receive element sets have an included angle to provide a pseudo-focusing effect in the inspected material.

Enquiry

Features

Temperature & Testing Data

This data is calculated with the transducer, wedge, and asset all at 150°C (peek data) or 200°C (celazole data). During actual inspections, there will be a temperature gradient from the asset up to the transducer.
 
 

High-Temp Wedges – Celazole U-60

Up to 200°C (392°F)
 

 

Peek
Temp (°C) 1st BW Gain (dB) Low / Low Velocity (in./uS) % Dif V from 25 °C Gain at 80% FSH % Dif BW from 25(°C) Fc (MHz) % Dif Fc from 25 (°C)
25 10 0.117 0 16 0 4.88 0
50 10 0.116 0.77 16 0 4.8 1.64
75 10 0.116 1.20< 16.5 3.13< 4.76 2.46>
100 12.5 0.115 1.97< 17.5 9.38 4.66 >4.51
120 15 0.114 2.65 19.5 21.88 4.27 12.50
150 20 0.113 3.68 23.5 46.88 3.86 20.90
175 26 0.113 3.42 29.5 84.38 3.91 19.88
200 39 0.112 4.53 39 143.75 6.05 -23.98

Specifications

Part Numbers
Case Style Frequency Number of Elements Element Pitch Elevation Part Number Replaceable Delay
inch mm inch mm
CL 5 64 (32 x 2) 0.058 1.5 0.20 5 00-015634 Celazole 01-014525

View the Technology & Features tab for more information including temperature and testing data.
 
 

Case Dimensions
Case Length Width Height A Height B
CL 2.58 in. 65.5 mm 1.25 in. 31.8 mm 0.98 in. 24.9 mm 0.22 in. 5.6 mm

 